3-50
Service Guide E8364-90001
Tests and Adjustments
PNA Series Microwave Network Analyzers
Adjustments
E8362A, E8363A, E8364A
3.8 GHz PMYO Frequency Adjustment
The 3.8 GHz adjustment is used to adjust the frequency accuracy of the network analyzer’s
3.8 GHz permanent-magnet YIG oscillator (PMYO) on the A12 source 20.
Equipment used for the Frequency Adjustment at 3.8 GHz
Procedure
1. Connect the equipment as shown in
. Connect a GPIB cable between the
network analyzer and the spectrum analyzer.
Figure 3-24 Setup for Adjustment of the 3.8 GHz PMYO
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
3.8 GHz Freq Adjust
.
3. Ensure the GPIB settings are correct.
4. Click
Begin Adj
, and then follow the instructions as they are displayed.
Equipment Type
Models
Used With
Model or
Part Number
Alternate Model
or Part Number
Spectrum analyzer
All
8565E
856xE
a
a. Must be capable of measuring a signal at 3.8 GHz.
RF Cable, 3.5 mm (f) to 3.5 mm (f)
E8362A
85131C
85131E
Adapter, 3. 5 mm (m) to type-N (m)
E8362A
1250-1743
None
RF Cable, 2.4 mm (f) to 2.4 mm (f)
E8363A, E8364A
85133C
85133E
Adapter, 2.4 mm (m) to type-N (m)
E8363A, E8364A
11903A
None
Summary of Contents for E8362A
Page 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Page 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Page 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Page 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Page 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Page 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Page 287: ...Service Guide E8364 90001 A 1 A Error Terms ...
Page 303: ...Service Guide E8364 90001 B 1 B Option Enable Utility ...
Page 309: ...Service Guide E8364 90001 C 1 C Firmware Upgrades ...
Page 313: ...Service Guide E8364 90001 D 1 D Operating System Recovery ...