1
1
Battery Management Functionality
E7495A/B Manual Verification
•
Verify that the Front Panel Battery LED’s accurately indicates
the correct status of the battery(s) installed in the Test Set.
Battery 1:
(Light Grey: No Battery Installed)
Battery 2
should be one of the
following:
(Green: Greater than 25% Charge remaining)
(Yellow: 10 to 15% Charge remaining)
(Red: Less than 10% Charge remaining)
(Dark Grey: No Charge remaining)
Battery 1:
(Off: No Battery Installed)
Battery 2
should be one of
the following:
(Blinking Green: Battery is Charging)
(Solid Green: Battery in Fully Charged)