24
Chapter 3
Operation Verification
Operation Verification
Operation Verification
Operation verification is a series of tests that, when completed, will either ensure that the signal generator is
operating correctly, or will assist in pointing to the problem area. Operation verification does not ensure
performance to specifications, but should provide a level of confidence that the signal generator is operating
correctly within a minimum amount of time.
Operation verification is appropriate for incoming inspection, after repair (when a full calibrated
performance is not required), or whenever the integrity of the signal generator is in question.
Perform the following tests in the order they are presented. The tables referenced by the tests are located in
the rear of the chapter, where they can be copied easily.
1.
Power On the Signal Generator on page 25
2.
Check for Error Messages on page 26
3.
Frequency Range and Accuracy Check on page 27
4.
Power Level Accuracy Check on page 29
5.
6.
7.
I/Q Modulation Check on page 37
8.
Option UN7 (BERT) Check on page 38
Required Equipment
•
Agilent 53132A Option 050 Frequency Counter
•
Agilent E4418B or E4419B Power Meter
•
Agilent E9304A Power Sensor
•
Agilent E444xA Spectrum Analyzers with Option 1DS
•
Agilent 8491A/B Option 006 Attenuator (6 dB)
•
Agilent 8491A/B Option 010 Attenuator (10 dB)
•
Agilent 8902A Measuring Receiver
Summary of Contents for E4428C ESG
Page 7: ...1 1 Safety Information ...
Page 12: ...6 Chapter 1 Safety Information General Safety Considerations ...
Page 13: ...7 2 Getting Started ...
Page 28: ...22 Chapter 2 Getting Started Contacting Agilent Sales and Service Offices ...
Page 29: ...23 3 Operation Verification ...
Page 61: ...55 4 Regulatory Information ...
Page 64: ...58 Chapter 4 Regulatory Information Regulatory Information misc ...