Frequency Range and Accuracy Check
NOTE
For greater counter resolution accuracy, set the gate time to > 5 seconds for maximum
counter accuracy. Press
Gate & ExtArm
twice and use the arrow keys to set value.
Measure the Frequency Accuracy
1. Set the signal generator to each frequency listed in
Table 3-7, “Frequency Accuracy,” on page 44
2. Record the measured frequency in
and compare it to the corresponding limits.