26
Chapter 3
Operation Verification
Check for Error Messages
Check for Error Messages
1. Check the display to see if the
ERR
annunciator is turned on.
2. If the
ERR
annunciator is turned on, review the error messages in the queue by pressing
Utility
>
Error Info
. The first error message in the queue will be shown in the text area of the display.
Refer to the “Error Messages” on the Documentation
CD-ROM
for further information about specific
error messages. Refer to
Table 2-2, “Available ESG Documentation,” on page 17
If there is more than one error message (each message will be designated as 1 of
n
), press the
View Next Error Message
softkey until you have seen all of the messages.
3. When you have resolved all of the error messages, press
Clear Error Queue(s)
to delete the messages.
4. Cycle the power on the signal generator and then restart this procedure until the signal generator powers
on without displaying the
ERR
annunciator.