Operation Verification
I/Q Modulation Check
I/Q Modulation Check
NOTE
The I/Q modulation check is available only in the E4438C ESG vector signal generator with
Option 001/601 or 002/602.
1. Press
Preset
.
2. Press
I/Q
>
I/Q Calibration > Calibration Type DC User Full
until Full is highlighted.
3. Press
Execute Cal
to begin an I/Q calibration of the full frequency range of the signal generator.
The signal generator displays the message
I/Q Calibration in Progress
while the calibration is
occurring.
To abort the I/Q calibration, press
Abort Cal
. The message
I/Q Calibration in Progress
Aborting...
is displayed until the calibration is stopped completely.
NOTE
The I/Q calibration is stored in non-volatile memory and remains unchanged during a preset
or the power cycle of the signal generator.