Agilent B2200 User’s Guide, Edition 2
3-15
Front Panel Operation
Switch Control Functions
Connection Sequence
NOTE
This is only for modules that are set to the single connection rule. Refer to
“Connection Rule” on page 3-14.
Connection sequence specifies the open/close sequence of the relays when changing
from an existing connection to a new connection.
You can select one of the three connection sequences shown in Table 3-1 for each
module installed in the Agilent B2200.
When an existing connection is changed to a new connection, the relays are
opened/closed as follows:
Table 3-1
Connection Sequence
Connection
Sequence
Relay Operation
Break Before Make
1. Disconnect previous route.
2. Wait for relays to open.
3. Connect new route.
Make Before Break
1. Connect new route.
2. Wait for relays to close.
3. Disconnect previous route.
No Sequence
1. Disconnect previous route.
2. Connect new route.
Summary of Contents for B2200A
Page 17: ...1 Introduction ...
Page 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Page 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Page 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Page 35: ...2 Installation ...
Page 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Page 61: ...3 Front Panel Operation ...
Page 99: ...4 Programming ...
Page 127: ...5 SCPI Command Reference ...
Page 187: ...6 VXI plug play Driver ...
Page 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Page 221: ...7 Error Messages ...