5-8
Agilent B2200 User’s Guide, Edition 2
SCPI Command Reference
*OPC
*OPC
This command starts to monitor pending operations, and sets/clears the Operation
Complete (OPC) bit in the Standard Event Status Register as follows:
•
If there is no pending operation, sets the OPC bit to 1.
•
If there are any pending operations, sets the OPC bit to 0. The bit will be set to 1
again when all pending operations are completed.
So, *OPC command is required to enable the OPC bit. To stop monitoring pending
operations (disable OPC bit), execute the *CLS command.
For details about the Standard Event Status Register, see “Standard Event Status
Register” on page 5-58. Also, see *WAI command.
Syntax
*OPC
Query response
1 <newline><^END>
*OPC?
places ASCII character
1
into the Output Queue when all pending
operations are completed. For details, see “Output Queue” on page 5-60.
Example
OUTPUT @Agb2200;"*OPC"
The following example is for query:
OUTPUT @Agb2200;"*OPC?"
ENTER @Agb2200;A
Summary of Contents for B2200A
Page 17: ...1 Introduction ...
Page 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Page 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Page 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Page 35: ...2 Installation ...
Page 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Page 61: ...3 Front Panel Operation ...
Page 99: ...4 Programming ...
Page 127: ...5 SCPI Command Reference ...
Page 187: ...6 VXI plug play Driver ...
Page 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Page 221: ...7 Error Messages ...