Agilent B2200 User’s Guide, Edition 2
Front Panel Operation
Switch Control Functions
Connection Rule
Connection rule is used to specify if an input or output port can have multiple
connections. For each module installed in the mainframe (for auto configuration
mode, installed modules are treated as one module), you can specify one of the
following connection rules:
Single:
Each input port can be connected to
only one
output port, and
each output port can be connected to
only one
input port. So,
existing connection to a port will be disconnected when a new
connection is made.
If normal configuration mode is set to the Agilent B2200 with
multiple modules, an input/output path is effective for each
module. For example, if normal configuration mode is set to the
Agilent B2200 with two modules, and single rule is set to both
modules, the following connection is possible.
Free:
Each input port can be connected to
multiple
output ports, and
each output port can be connected to
multiple
input ports.
CAUTION
If the Free connection rule has been specified, ensure multiple input ports are not
connected to the same output port. Such configurations can cause damage.
Module 1 Output
1
12
2
1
Close
Open
14
3
2
Module 2 Output
1
2
12
3
3
Input
Summary of Contents for B2200A
Page 17: ...1 Introduction ...
Page 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Page 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Page 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Page 35: ...2 Installation ...
Page 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Page 61: ...3 Front Panel Operation ...
Page 99: ...4 Programming ...
Page 127: ...5 SCPI Command Reference ...
Page 187: ...6 VXI plug play Driver ...
Page 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Page 221: ...7 Error Messages ...