18
User Manual
2
Setup
Switches 1 and 2 affect the beeper operation.
The remaining four switches combine to specify either a test mode or normal
operation for each channel. The test mode provides service technicians with a
diagnostic tool to show the effects of electrical noise on a known test signal.
Switches 3 and 4 combine to specify either a test mode or normal operation
mode for Signal A.
Switches 5 and 6 combine to specify either a test mode or normal operation
mode for Signal B.
Table 1
Configuration switch settings
Switch
Function
Position
Effect
1
Beep/Off
Up
Causes an audible beep when Start/Stop is pressed.
Down
Silent at Start/Stop.
2
HI/LO
Up
Beep, if any, is loud.
Down
Beep, if any, is soft.
3
Channel A -
TEST/NORM
Up
Selects stored test chromatogram. Analog input signal ignored.
Down
Selects Analog input for Normal Operation.
4
Channel A -
ADD/IGNORE
Up
Add signal input to the stored chromatogram. Switch 3 must be in Up
position.
Down
Do not add (ignore) signal input to stored chromatogram for Channel A.
5
Channel B -
TEST/NORM
Up
Selects stored test chromatogram. Analog input signal ignored.
Down
Selects Analog input for Normal Operation.
6
Channel B-
ADD/IGNORE
Up
Add Channel B signal input to the stored chromatogram. Switch 5 must be
in Up position.
Down
Do not add (ignore) signal input to stored chromatogram for Channel B.
Summary of Contents for Agilent 35900E
Page 1: ...Agilent Technologies Agilent 35900E Dual Channel Interface User Manual ...
Page 6: ...6 User Information Manual ...
Page 36: ...36 User Manual 4 Instrument Connection ...
Page 48: ...48 User Manual 6 Parts Replacement ...
Page 51: ...Instrument Parts Breakdown 7 User Manual 51 Figure 15 Parts breakdown 1 2 3 4 5 ...
Page 54: ...54 User Manual 7 Instrument Parts Breakdown ...
Page 60: ...60 User Manual A Cables ...