24
User Manual
3
Computer Connection
Configuration for RS-232 operation
The GPIB/RS-232 MIO card contains a single 8 position configuration switch
(Figure 6). When the position 1 switch is set to ON, the MIO card uses the
RS-232 communication mode. In this mode, the ON switch positions for
switches 2 through 8 are defined in Table 3.
An example setting for this switch is shown in Figure 6.
Table 3
RS-232 communication mode
Switch
ON Action
2
Handshake enabled
3
Hardware handshake (otherwise software handshake)
4
Parity enable
5
Even parity (otherwise odd parity)
6, 7, 8
Baud rate (see Table 4)
Figure 6
Example settings for RS-232
1
8
On
Off
Rear
Summary of Contents for Agilent 35900E
Page 1: ...Agilent Technologies Agilent 35900E Dual Channel Interface User Manual ...
Page 6: ...6 User Information Manual ...
Page 36: ...36 User Manual 4 Instrument Connection ...
Page 48: ...48 User Manual 6 Parts Replacement ...
Page 51: ...Instrument Parts Breakdown 7 User Manual 51 Figure 15 Parts breakdown 1 2 3 4 5 ...
Page 54: ...54 User Manual 7 Instrument Parts Breakdown ...
Page 60: ...60 User Manual A Cables ...