This chapter provides basic instructions to isolate a problem.
Troubleshooting Hints
This section provides a brief checklist of common failures. Before troubleshooting the
instrument, be sure the failure is in the instrument rather than any external connections.
Unit seems inoperative
1
Verify the ac power cord is connected to the instrument.
2
Press the front panel power switch.
3
Listen for the fan (on the right side of the instrument) to isolate a problem with the
front panel.
4
Remove all plug-in modules to verify a module is not causing the failure.
5
Verify the power supply as described on
Isolating to an assembly
1
Remove all plug-in modules from the mainframe.
2
Remove any connections to the Analog Bus.
3
Turn on power and listen for the beep and fan operation. The fan operates from the
main power supply.
4
Isolate the internal DMM by removing it from the instrument. The instrument should
turn on and pass self-test without the internal DMM installed. Disassembly
procedures begin on
.
Unit fails self-test
1
Check the display for the power on self test results.
2
Remove all plug-in modules from the mainframe and any connections to the Analog
Bus. Run the self-test again. A passing self-test indicates a problem with a plug-in
module. Replace the plug-in modules one at a time to isolate the failing module.
3
If the self test still fails, remove the internal DMM from the instrument. The
instrument should turn on and pass self-test without the internal DMM installed.
Disassembly procedures begin on
4
Self-test errors are described beginning on
深圳德标仪器
135-1095-0799
Summary of Contents for Agilent 34980A
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...iii 深圳德标仪器 135 1095 0799 ...
Page 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Page 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Page 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Page 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Page 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...