Calibration Procedures
3
34980A Service Guide
51
Calibration Process
The following general procedure is the recommended method to complete a full
instrument calibration.
1
on page 55 and
2
Perform the verification tests to characterize the instrument (incoming data).
3
Unsecure the instrument for calibration (
4
Perform the DMM adjustment procedures (
5
Perform the DAC adjustment procedures if either the 34951A (
on page 68) or 34952A (
6
Secure the instrument against calibration.
7
Note the new security code and calibration count in the instrument’s maintenance
records.
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration after the procedure has already
been initiated. You can abort a calibration at any time on any module by turning off the
power. You can abort a calibration on the internal DMM or the 34951A 4-Ch Isolated
DAC by issuing a remote interface device clear message. You can abort a calibration on
the 34952A Multifunction Module by sending the
CALibration:ABORt
command.
NOTE
The 34951A 4-Ch Isolated DAC is intended to be adjusted frequently to
compensate for changes in the module’s environment. These
adjustments can be made either volatile or non-volatile. See
details.
CAUTION
If you abort a calibration in progress when the instrument is attempting
to write new calibration constants to EEPROM, you may lose all
calibration constants for the function. Typically, upon re-applying
power, the instrument will report error
705 Cal:Aborted
. You may also
generate errors 740 through 746. If this occurs, you should not use the
instrument until a complete re-adjustment has been performed. For a list
of calibration error numbers, see
.
深圳德标仪器
135-1095-0799
Summary of Contents for Agilent 34980A
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...iii 深圳德标仪器 135 1095 0799 ...
Page 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Page 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Page 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Page 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Page 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...