74
34980A Service Guide
3
Calibration Procedures
Analog Output Adjustment
Install the 34952A module in the mainframe and allow a
1 hour warm-up before performing these procedures.
This adjustment procedure sets a zero adjustment and a gain adjustment constant for
each DAC output. You must perform all the adjustments on one analog output channel
before adjusting the other analog output channel.
1
Install the module(s) in the instrument.
2
Unsecure the instrument for calibration (see
3
Connect channel 6 DAC output to the DMM input. Set the DMM to measure DC
volts.
4
The calibration procedure makes two adjustments per channel.
After sending the first command, measure the module output. Send the measured
value to the module and advance to the next point.
This procedure is summarized as follows:
a
Send the following command to begin the procedure.
CALibration:BEGin:VOLTage 1,(@<channel>
)
b
Measure the module output.
c
Send the measured value to the module with the following command:
CALibration:POINt? <value>
d
The command returns a “+1” to indicate it is ready for the next point.
e
Measure the module output.
f
Send the measured value to the module with the following command:
CALibration:POINt? <value>
g
The command returns a “+0” to indicate the calibration on the channel is
completed.
5
Repeat steps 3 and 4 for channel 7.
Output Voltage
Quick Check Error
from
Nominal (1 year)
10 V
Q
± 45 mV
0 V
Q
± 20 mV
-10 V
± 45 mV
深圳德标仪器
135-1095-0799
Summary of Contents for Agilent 34980A
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...iii 深圳德标仪器 135 1095 0799 ...
Page 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Page 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Page 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Page 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Page 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...