12
34980A Service Guide
2
Specifications
Multiplexer Module Specifications and Characteristics
[1]
DC or AC RMS voltage, channel-to-channel or channel-to-earth
[2]
Peak voltage, channel-to-channel or channel-to-earth
[3]
Into analog bus. System errors are included in the internal DMM measurement accuracy specifications.
[4]
With in-rush resistors bypassed. Bypassing resistors reduces lifetime of relays. See the rated load relay life characteristics.
[5]
Limited to 6 W of channel resistance power loss per module
[6]
DC or peak AC current
[7]
Ambient temperature < 30
°
C
[8]
Includes 0.5
°
C temperature reference sensor and 0.5
°
C terminal block isothermal gradient error. Measured under worst case
loading of the mainframe. See 34980A User's Guide for information on supported external reference sensors.
[9]
With 100
input protection resistors.
34921A
34922A
34923A
34924A
34925A
Channels/configurations
40 2-wire
20 4-wire
4-current
1.5 A Fused
70 2-wire
35 4-wire
80 1-wire
40 2-wire
20 4-wire
70 2-wire
35 4-wire
80 1-wire
40 2-wire
20 4-wire
Switch type
Armature
latching
Armature
latching
Reed
Reed
Optically
isolated FET
Input characteristics (per channel)
Max volts
± 300 V
[1]
± 300 V
[1]
± 150 V peak
[2]
± 150 V peak
[2]
± 80 V peak
[2]
Max current (DC, AC RMS)
Switch current
Carry current
1 A
2 A
1 A
2 A
0.5 A
[4]
/0.05 A
[9]
1.5 A
[4]
/0.05 A
[9]
0.5 A
[5]
/0.05 A
[9]
1.5 A
[5]
/0.05 A
[9]
0.02 A
[6]
Power (W, VA)
[5]
60 W
60 W
10 W
10 W
1.6 W
Volt-Hertz limit
10
8
10
8
10
8
10
8
10
7
Initial closed channel res
[3][10]
< 1.5
< 1.5
< 1.5
5
/200
9
< 1.5
5
/200
9
< 700
General specifications
Offset voltage
[3]
< 3
V
< 3
V
< 50
V
< 100
V 1-wire
< 50
V
< 3
V
DC Isolation (ch-ch, ch-earth) >10 G
>10 G
>10 G
>10 G
>10 G
Leakage current
[3]
N/A
N/A
N/A
N/A
20 nA
[7]
T/C cold junction accuracy
[3, 8]
< 1 °C
N/A
N/A
N/A
N/A
深圳德标仪器
135-1095-0799
Summary of Contents for Agilent 34980A
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...iii 深圳德标仪器 135 1095 0799 ...
Page 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Page 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Page 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Page 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Page 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...