62
34980A Service Guide
3
Calibration Procedures
Internal DMM Adjustments
You will need a test input fixture to adjust the internal DMM
Zero Adjustment
Each time you perform a zero adjustment, the Internal DMM stores a new set of offset
correction constants for every measurement function and range. The Internal DMM will
sequence through all required functions and ranges automatically and store new zero
offset calibration constants. All offset corrections are determined automatically.
You may
not correct a single range or function without re-entering ALL zero offset correction
constants automatically. This feature is intended to save calibration time and improve
zero calibration consistency.
Zero Adjustment Procedure
The zero adjustment procedure takes about 5 minutes to complete. Be sure to allow the
instrument to warm up for 2 hours before performing the adjustments.
Follow the steps outlined below. Review
on page 55 before
beginning this test.
1
This procedure will use the copper shorts installed on input test connector. Leave the
Current input connection open.
2
Set the DC VOLTS function.
3
Send the value 0.000000 to the instrument using the
CALibration:VALue
0.000000
command.
4
Calibrate the instrument using the
CALibration?
command.
5
Perform the
on page 55 to check zero calibration results.
NOTE
Never turn off the Internal DMM during Zero Adjustment. This may
cause ALL calibration memory to be lost.
深圳德标仪器
135-1095-0799
Summary of Contents for Agilent 34980A
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...iii 深圳德标仪器 135 1095 0799 ...
Page 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Page 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Page 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Page 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Page 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...