Troubleshooting and Diagnostics
4
34980A Service Guide
81
Power Supply
The main power supply pr12 Volts. All other power supplies are derived from
this supply. To test the main power supply:
1
Disassemble the instrument as described beginning on
2
Use a DVM to check that the power supply output is +12 V ± 0.6 V.
Test the power supply output at the red leads on the power supply connector to the
communications board.
a
The +12 Volt supply is always on when the ac power is applied to the instrument.
Test for loading errors by unplugging the power supply connector from the
communications board.
3
Use an oscilloscope to verify there is no ac oscillation on the +12 V power supply.
Replace the power supply assembly if +12 V is not present.
Product Firmware Updates
As new product features and enhancements become available, you can easily update
your mainframe and plug-in module firmware to ensure optimum compatibility. The
latest firmware updates are available from the Agilent 34980A product page at
www.agilent.com/find/34980AUpdates
Front Panel Operation:
Utility > FIRMWARE > UPDATE
Once you have downloaded the latest mainframe firmware (see above),
use the knob to scroll through the installed modules that require a firmware update. To
exit the menu without installing the updates,
select CANCEL.
WARNING
Exposed Mains. When the instrument cover is removed to test
the power supply, the ac mains are exposed.
深圳德标仪器
135-1095-0799
Summary of Contents for Agilent 34980A
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...iii 深圳德标仪器 135 1095 0799 ...
Page 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Page 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Page 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Page 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Page 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...