70
Chapter 3
Troubleshooting
AF, RF, & CDMA Diagnostics (Step 3)
RF Diagnostics
This program tests the RF functions of the following assemblies:
• A2A115 Downconverter
• A2A24 RF Output
• A2A25 Signal Generator Synthesizer
• A2A23 Reference
• A2A21 Receiver
• A2A22 Receiver Synthesizer
• A2A20 Spectrum Analyzer
• A2A130 RF I/O
• A2A110 Upconverter
Some tests require cabling before the RF Diagnostics can be run; but all
tests can be run in a loop mode without further intervention. Running
in loop mode makes it easier to catch intermittent failures. To run these
diagnostics, see
“RF Diagnostics” on page 70
.
NOTE
A fifteen minute warm up is required. The measurement limits of the
SERVICE4 diagnostic tests are valid only at room temperature; that is,
20º to 25ºC (65º to 75ºF).
Summary of Contents for 8935 Series
Page 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Page 93: ...92 Chapter 3 Troubleshooting Product Verification...
Page 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Page 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Page 157: ...156 Chapter 6 Replaceable Parts Parts List...
Page 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Page 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Page 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Page 322: ...321 A Error Messages...
Page 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...