![Agilent Technologies 8935 Series Manual Download Page 193](http://html1.mh-extra.com/html/agilent-technologies/8935-series/8935-series_manual_2868625193.webp)
192
Chapter 8
Performance Tests
RF Generator Level Accuracy Performance Test 5
Figure 8-10
Setup 2 for Measurements of 1700 and 2000 MHz
Procedure 2
Steps 1 to 5 in the following procedure apply to Setup 1 shown in
Figure
8-9 on page 190
.
1. Connect the sensor module on the measuring receiver to the
DUPLEX OUT port of the Test Set.
2. On the Test Set:
a. Set the
Amplitude
to
-10 dBm
.
b. Set the
RF Gen Freq
to
1700 MHz
.
3. On the measuring receiver:
a. Set the measurement mode to RF power.
b. Key in 1700 MHz.
c. Measure and record the RF power at the DUPLEX OUT port.
4. On the Test Set set the RF Gen Freq to 2000 MHz.
5. On the measuring receiver:
a. Key in 2000 MHz.
b. Measure and record the RF power at the DUPLEX OUT port.
Steps 6 to 8 apply to Setup 2 shown in
Figure 8-10 on page 192
.
6. Make the connections shown in Setup 2.
DUPLEX OUT
Measuring Receiver
INPUT 50
Ω
IF OUTPUT
LO
RF
RF
INPUT
INPUT
OUTPUT
Signal Generator
Microwave
Converter
INPUT
OUTPUT
AMPLIFIER 1
RF IN/OUT
Summary of Contents for 8935 Series
Page 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Page 93: ...92 Chapter 3 Troubleshooting Product Verification...
Page 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Page 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Page 157: ...156 Chapter 6 Replaceable Parts Parts List...
Page 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Page 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Page 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Page 322: ...321 A Error Messages...
Page 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...