Chapter 10
289
Block Diagrams
RF Analyzer Section
Spectrum Analysis
The LO on the A2A20 Spectrum Analyzer is swept across the span by
the Controller, see
Figure 10-6 on page 293
. The LO starts sweeping
when the oscilloscope circuits on the A2A33 Measurement board trigger
the display sweep to start. As the LO sweeps, the spectrum analyzer
filters and then amplifies the IF signal in a logarithmic detector so the
signal voltage will be proportional to the log of power. The signal
voltage is measured by a sampler on the Measurement board and
displayed.
Spectrum analyzer resolution bandwidth is determined by switching
bandwidth IF filters on the A2A20 Spectrum Analyzer. These filters are
set by the Controller as a function of the span selected from the front
panel.
Summary of Contents for 8935 Series
Page 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Page 93: ...92 Chapter 3 Troubleshooting Product Verification...
Page 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Page 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Page 157: ...156 Chapter 6 Replaceable Parts Parts List...
Page 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Page 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Page 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Page 322: ...321 A Error Messages...
Page 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...