![Agilent Technologies 8935 Series Manual Download Page 195](http://html1.mh-extra.com/html/agilent-technologies/8935-series/8935-series_manual_2868625195.webp)
194
Chapter 8
Performance Tests
RF Generator Level Accuracy Performance Test 5
+ RF level measured at -80 dBm in step 8i
– RF level measured at -80 dBm in step 8j
+ RF level measured in step 8k
For example, if:
RF power measured at
−
10 dBm in step 3b or 5b =
−
10.2 dBm
RF level measured at
−
80 dBm in step 8i =
−
70.1 dB
RF level measured at -80 dBm in step 8j =
−
52.6 dB
RF level measured at -100 dBm in step 8k =
−
73.2 dB the
corrected level at
−
100 dBm is
−
10.2 + (
−
70.1)
−
(
−
52.6) + (
−
73.2) =
100.9 dBm. Compare the corrected values with the limits in the
PTR.