Chapter 8
195
Performance Tests
RF Generator Harmonics Spectral Purity Performance Test 6
RF Generator Harmonics Spectral Purity
Performance Test 6
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record)
Table 9-6, “RF
Generator Harmonics Spectral Purity Test 6 Record,” on page 250
.
Harmonic signals with the carrier set to several frequencies and two
different levels (maximum output and minimum level vernier) are
searched for by an RF spectrum analyzer.
Figure 8-11
Setup
Procedure
1. Set up the spectrum analyzer in accordance with its operating
manual.
2. On the Test Set:
a. Press
Preset
.
b. Select the INSTRUMENT CONFIGURE screen.
c. Set the
RF Display
field to
Freq
.
d. Select the CDMA GENERATOR screen.
e. Set the
CW RF Path
field to
Bypass
.
f. Select the RF GENERATOR screen.
g. Set the
RF Gen Freq
to
1 MHz
.
h. Set the
Amplitude
to
-10 dBm
.
3. Set the test set’s RF generator to the frequencies and levels shown in
the Performance Test Record (PTR) and measure the second and
third harmonics. For each measurement convert the harmonic level
to dB below the fundamental (dBc) and compare the computed levels
to the limits.
INPUT 50
Ω
DUPLEX OUT
Spectrum Analyzer
Summary of Contents for 8935 Series
Page 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Page 93: ...92 Chapter 3 Troubleshooting Product Verification...
Page 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Page 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Page 157: ...156 Chapter 6 Replaceable Parts Parts List...
Page 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Page 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Page 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Page 322: ...321 A Error Messages...
Page 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...