![Agilent Technologies 8935 Series Manual Download Page 229](http://html1.mh-extra.com/html/agilent-technologies/8935-series/8935-series_manual_2868625229.webp)
228
Chapter 8
Performance Tests
CDMA Analyzer Channel Power Level Accuracy Performance Test 28
2. On the signal generator:
a. Set the frequency to 881.32 MHz.
b. Set the output level so the measuring receiver reads 11 dBm
3. Push
Preset
on the Test Set.
4. On the Test Set:
a. Change the
Avg Pwr
field to
Chan Pwr
.
b. Set the
Tune Freq
to 881.32 MHz
c. Select Calibrate under
Chn Pwr Cal
.
5. Record the
Chan Pwr
measurement in the PTR, see
Table 9-29,
“CDMA Analyzer Channel Power Level Accuracy Test 28 Record,” on
page 280
.
6. Repeat steps 2 and 3 for each of the data points listed in the PTR.
NOTE
The
Chan Pwr Cal
is only required when the frequency is changed.