2-6
Chapter 2
Introduction to Instrument Control
Instrument Control using the VXIplug&play Driver
Interface Bus Verification Procedure (National Instruments Card)
1. Check the bus connection by running Win32 VISA Interactive Control. When this
program is run, it automatically reports what it finds on the bus. Refer to
Notice that the program is reporting instruments at addresses 16 and 17. As mentioned
earlier, these addresses designate the instrument and its display, respectively.
Figure 2-2 Win32 VISA Interactive Control Window: Bus Report
2. To further verify GPIB operation, double click on “GPIB0::16::INSTR” and then perform
the following steps. Refer to
for a view of the window you will be using.
a. Make sure that the “Basic I/O” tab is selected.
b. Click on the “Write” tab.
c. Enter
PRES;
in the “Buffer” text box.
d. Click on “Execute.”
e. This command should preset the analyzer. If an instrument preset does not occur,
there is a problem. Check all GPIB address settings and physical connections. Most
GPIB problems are caused by an incorrect address or faulty/loose GPIB cables.
Summary of Contents for 8719ES
Page 15: ...1 1 1 Alphabetical Command Reference ...
Page 293: ...2 1 2 Introduction to Instrument Control ...
Page 310: ...3 1 3 GPIB Programming ...
Page 334: ...4 1 4 Reading Analyzer Data ...
Page 343: ...5 1 5 Data Processing Chain ...
Page 350: ...6 1 6 Error Reporting ...
Page 364: ...7 1 7 Programming Examples ...
Page 502: ...A 1 A Preset Conditions ...
Page 517: ...B 1 B Command Listings ...