![Agilent Technologies 8719ES Programmer'S Manual Download Page 282](http://html.mh-extra.com/html/agilent-technologies/8719es/8719es_programmers-manual_2867960282.webp)
1-268
Chapter 1
Alphabetical Command Reference
TSTIO
TSTIO
Syntax
TSTIO<FWD|REV><num>;
or
TSTIO<FWD|REV>?;
Description
Both of these commands define 3 bits, D0 through D2, on the test set connector I/O. Be
careful that you do not also set a value to
ATTP1
and
ATTP2
as there is interaction between
these commands and they will change the values you have set for D0 through D2 and will
couple the channels together. Values for
ATTP1
and
ATTP2
translate to the following values
for D0 through D2:
Front Panel Equivalents
ATTP1/ATTP2
D0–D2
0 dB
7
10 dB
6
20 dB
5
30 dB
4
40 dB
3
50 dB
2
60 dB
1
70 dB
0
Command
Description
Range
Query Response
TSTIOFWD
Defines D0–D2 on the test set connector
I/O for chan 1 and chan 2 forward setting.
integers 0–7
<num><
L
F
>
TSTIOREV
Defines D0–D2 on the test set connector
I/O for chan 1 and chan 2 reverse setting.
integers 0–7
<num><
L
F
>
Command
Hardkey
Softkey
TSTIOFWD
TSTIOREV
Seq
TESTSET I/O REV
Seq
TESTSET I/O FWD
Summary of Contents for 8719ES
Page 15: ...1 1 1 Alphabetical Command Reference ...
Page 293: ...2 1 2 Introduction to Instrument Control ...
Page 310: ...3 1 3 GPIB Programming ...
Page 334: ...4 1 4 Reading Analyzer Data ...
Page 343: ...5 1 5 Data Processing Chain ...
Page 350: ...6 1 6 Error Reporting ...
Page 364: ...7 1 7 Programming Examples ...
Page 502: ...A 1 A Preset Conditions ...
Page 517: ...B 1 B Command Listings ...