Chapter 1
1-157
Alphabetical Command Reference
OUTP
OUTPLIML
Outputs the limit test results for each
point in the sweep. This is an ASCII
transfer.
N/A
<array><
L
F
>
OUTPLIMM
Outputs the limit test results at the active
marker.
N/A
<num,num,num,num>
<
L
F
>
OUTPMARK
Outputs the active marker values. The
first two numbers are the marker
response values, and the last is the
stimulus value.
See
meaning of the response values as a
function of display format.
N/A
<num,num,num><
L
F
>
OUTPMEMF
Fast data transfer command for
OUTPMEMO
N/A
<array><
L
F
>
OUTPMEMO
Outputs the memory trace from the active
channel. The data is in real/imaginary
pairs, and can be treated the same as data
read with the OUTPDATA command. See
N/A
<array><
L
F
>
OUTPMSTA
Outputs the marker statistics in ASCII
format: mean, standard deviation, and
peak-to-peak variation in that order. If
statistics is not on, it is turned on to
generate current values and turned off
again. See also
N/A
<num,num,num><
L
F
>
OUTPMWID
Outputs the marker bandwidths search
results in ASCII format: bandwidth,
center, and Q in that order. If widths is not
on, it is turned on to generate current
values and then turned off again.
N/A
<num,num,num><
L
F
>
OUTPMWIL
Outputs the marker bandwidths search
results in ASCII format: bandwidth,
center, Q, and loss in that order. If widths
is not on, it is turned on to generate
current values and turned off again.
N/A
<num,num,num,num>
<
L
F
>
OUTPOPTS
Outputs an ASCII string of the options
installed in the analyzer.
N/A
<$><
L
F
>
OUTPPLOT
Outputs the HP-GL plot string in ASCII
format to the GPIB port. Can be directed
to a plotter, or read into the computer.
N/A
<$><
L
F
>
OUTPPMCAL
Outputs the power meter calibration array
for channel 1 or channel 2. A default array
is used if a power meter calibration sweep,
TAKCS
, has not been performed. Values are
returned as 100 times the power meter
reading in dB. This is an ASCII transfer
(FORM4).
integers 1 or 2
<array><
L
F
>
Command
Description
Range
Response
Summary of Contents for 8719ES
Page 15: ...1 1 1 Alphabetical Command Reference ...
Page 293: ...2 1 2 Introduction to Instrument Control ...
Page 310: ...3 1 3 GPIB Programming ...
Page 334: ...4 1 4 Reading Analyzer Data ...
Page 343: ...5 1 5 Data Processing Chain ...
Page 350: ...6 1 6 Error Reporting ...
Page 364: ...7 1 7 Programming Examples ...
Page 502: ...A 1 A Preset Conditions ...
Page 517: ...B 1 B Command Listings ...