![Agilent Technologies 85024A User'S And Service Manual Download Page 57](http://html.mh-extra.com/html/agilent-technologies/85024a/85024a_users-and-service-manual_2867925057.webp)
Chapter 7
7-11
Service
Replacement Procedure
Replacing the Type-N Output Connector
Tools Required:
9/16 inch open-end wrench
1. Remove the connector body with the open-end wrench.
2. Remove the male center conductor assembly from the gold plated inner connector. Use your fingers
to pull out the assembly. The small insulator may or may not come out attached to the center
conductor assembly. If it does, remove it from the old center conductor assembly and place it on the
new one. If it is not attached to the center conductor assembly (it remained inside the inner
connector), proceed to step 3.
3. Hold the new center conductor assembly by the large round insulator, not by the gold plated center
conductor. Carefully align the new center conductor and insert it into the inner connector. Press
gently until you feel it click into place.
4. Place the new connector body and tighten with the open-end wrench.
Figure 7-9
Output Connector Parts
NOTE
“Replaceable Parts” on page 6-4
for part numbers.
Summary of Contents for 85024A
Page 4: ...iv ...
Page 7: ...1 1 1 General Information ...
Page 13: ...2 1 2 Accessories ...
Page 19: ...3 1 3 Installation ...
Page 24: ...3 6 Chapter3 Installation Returning the Product for Service ...
Page 25: ...4 1 4 Operation ...
Page 30: ...4 6 Chapter4 Operation Operator s Check ...
Page 31: ...5 1 5 Performance Tests ...
Page 40: ...5 10 Chapter5 Performance Tests Average Noise Level ...
Page 41: ...6 1 6 Replaceable Parts ...
Page 46: ...6 6 Chapter6 Replaceable Parts Parts Lists ...
Page 47: ...7 1 7 Service ...
Page 60: ...7 14 Chapter7 Service Replacement Procedure Figure 7 11 Regulator Parts and Wiring ...
Page 64: ...7 18 Chapter7 Service Connector Inspection and Cleaning ...