![Agilent Technologies 85024A User'S And Service Manual Download Page 18](http://html.mh-extra.com/html/agilent-technologies/85024a/85024a_users-and-service-manual_2867925018.webp)
2-6
Chapter 2
Accessories
10:1 Divider
Replaceable Parts
If the tip is discolored, bent or broken, it must be replaced. Follow the procedure outlined in
below.
Figure 2-2
Exploded View of 10:1 Divider
Replacing the Probe Tip
1. Remove the 10:1 divider from the probe.
2. Unscrew the damaged tip with a 3/32 nut driver supplied with the probe and discard the tip.
3. Screw on the new tip and lightly tighten it with a 3/32 inch nut driver.
(Overtightening the tip can damage the nose assembly.)
Item
Description
Part Number
1
Short probe tip
85024-60015
Long probe tip
85024-60016
2
Guide
11881-20007