Performance Verification
7
SCD and NCD User Manual
107
Prepare for Chromatographic Checkout
Because of the differences in chromatographic performance
associated with different consumables, Agilent strongly
recommends using the parts listed here for all checkout tests.
Agilent also recommends installing new consumable parts
whenever the quality of the installed ones is not known. For
example, installing a new liner and septum ensures that they
will not contribute any contamination to the results.
When the GC is delivered from the factory, these consumable
parts are new and do not need replacement.
1
Check the indicators/dates on the gas supply traps. Replace
expended traps.
2
Install new consumable parts for the inlet and prepare the
correct injector syringe (and needle, as needed).
NOTE
For a new GC, check the installed inlet liner. The liner shipped in the inlet
may not be the liner recommended for checkout.
Table 13
Recommended parts for checkout by inlet type
Recommended part for checkout
Part number
Split splitless inlet
Syringe, 10-µL
5181-1267
Liner O-ring
5188-5365
Septum
5183-4757
Liner
5190-2295
Gold plated inlet seal, with washer
5188-5367
Multimode inlet
Syringe, 10-µL
5181-1267
Liner O-ring
5188-5365
Septum
5183-4757
Liner
5190-2295
Cool On-column inlet
Septum
5183-4758
Septum nut
19245-80521