Test Record Sheets
Performance Tests
Agilent 81630B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power Sensor Module, Second Edition
53
Performance Test for the Agilent 81630B
Page 3 of 3
Model Agilent 81630B Sensor Module
Report No.________
Date ________________
Test
No.
Test Description
Min. Spec.
Result
Max. Spec.
Measurement Un-
certainty
I
Accuracy Test
[
µ
W]
measured at _______ nm
(1310nm) Output Power
9.64
µ
W
_______
10.36
µ
W
measured at _______ nm
(1550nm) Output Power
9.64
µ
W
_______
10.36
µ
W
II
Linearity Test
For Calculations you may want to use the appropriate sheet
Range
P
DUT
[dBm]
P
DUT
[dBm]
Loss [%]
+20
+10
________
_______
<± 1.24
1
%
+10
+10 ________
_______
<±
1.16
%
+10
+3 ________
_______
<±
1.16
%
0
+3
________
_______
<± 1.16 %
0
-7
________
_______
<± 1.16 %
-10
-7
________
_______
<± 1.16 %
-10
-17
________
_______
<± 1.17 %
-20*
-17*
Reference
0.0
<± 1.17 %
-20
-27
________
_______
<± 1.22 %
-30
-27
________
_______
<± 1.22 %
-30
-37
________
_______
<± 1.76%
-40
-37
________
_______
<± 1.76 %
-40
-47
________
_______
<± 7.17 %
III
Noise Test
[pW]
_______
< 1.2 nW
IV
Return Loss Test
[dB]
55 dB <
_______
(V)
Relative Uncertainty due to Polarization
[dB]
Maximum difference in interval
1520 nm - 1580 nm / 0.75 nm step;
∆
P
_______
< 0.01 dB
(VI)
Relative Uncertainty due to Interference
[dB]
Maximum difference/FSR in interval
1520 nm - 1580 nm / 10 pm step;
∆
P
_______
< 0.005 dB
Summary of Contents for 81630B