Agilent 81630B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power Sensor Module, Second Edition
69
Cleaning Instructions
The following Cleaning Instructions contain some general safety
precautions, which must be observed during all phases of cleaning.
Consult your specific optical device manuals or guides for full
information on safety matters.
Please try, whenever possible, to use physically contacting connectors,
and dry connections. Clean the connectors, interfaces, and bushings
carefully after use.
If you are unsure of the correct cleaning procedure for your optical device, we
recommend that you first try cleaning a dummy or test device.
Agilent Technologies assume no liability for the customer’s failure to
comply with these requirements.
Cleaning Instructions for this Instrument
The Cleaning Instructions apply to a number of different types of
Optical Equipment. If you must clean the Agilent 81633A/4A Power
Sensor Modules, please refer the module to the skilled personnel of
Agilent’s service team. “How to clean instruments with a recessed lens
interface” on page 81 is relevant to all other Power Sensor modules.
Safety Precautions
Please follow the following safety rules:
• Do not remove instrument covers when operating.
• Ensure that the instrument is switched off throughout the cleaning
procedures.
• Use of controls or adjustments or performance of procedures other
than those specified may result in hazardous radiation exposure.
• Make sure that you disable all sources when you are cleaning any
optical interfaces.