Performance Tests
Performance Test Procedures
46
Agilent 81630B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power Sensor Module, Second Edition
Relative Uncertainty due to Polarization
(Optional Test)
N O T E
The performance test "Relative Uncertainty due to Polarization" is
optional, since the polarization is given with the production of the unit
by mechanical and optical cavities and stays unchanged by normal use
of the sensor module.
Below, you will find the test setup to verify the relative uncertainty
due to polarization of the sensor module. Generally, during this
measurement-procedure the tunable laser source is swept through a
predefined wavelength range. After every wavelength step, a single
PDL-measurement is made, where the polarization controller
generates all different polarization states. The highest PDL value is
taken as the "relative uncertainty due to polarization". The low output
power path of the coupler is used to monitor the power stability of the
setup