Performance Tests
Performance Test Procedures
44
Agilent 81630B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power Sensor Module, Second Edition
9
Press the [Measure] softkey to start the measurement.
10
After the stability application has finished, press the [Analysis]
softkey
,
press the [more]
softkey
, and note [
∆
P
] as the noise value in the test record.
Return Loss Test
1
Make sure that all connectors are clean.
2
Set up the equipment as shown in Figure 6.
3
Press
[Preset]
on the mainframe.
Figure 6
Return Loss Measurement Setup
N O T E
To ensure traceability, use the 81610CC Reference Cable for
calibration measurements. Do
not
use the 81610CC Reference Cable
for measurements on a Device Under Test. Instead, use a measurement
patchcord.
C A U T I O N
It is important to maintain the quality of the straight connector end of
the 81610CC Reference Cable.
Never
add another connector to the
straight end of the 81610CC Reference Cable, since a physical
connection is made. When the straight end of the 81610CC Reference
Cable is connected to the 81634B Power Sensor module, no physical
connection is made so there should be no degradation of connector
quality.
4
Zero the Return Loss Module and the Power Sensor Module
5
Enable the internal laser source of the Return Loss Module and
allow 20 minutes for it nto stabilize.
6
At the Power Meter:
Mainframe 8163B or 8164B w/
81657A and 8163xA/B (DUT)
8163 B
81613A 8163xB
81635A
in
out
81610CC