58
Chapter 3
VEE Measurement Program Operation
Setup Menu
Perform the Amplitude Calibration for E4402B Spectrum Analyzer
This chapter describes how to perform the amplitude calibration for E4402B Spectrum
Analyzer.
NOTE
The amplitude calibration must be performed at least once after the installation. Or the
error message will be displayed in the Narrow Band Spectrum measurements, such as
Narrow Band TAA, Overwrite, and Narrow Band Track Profile, etc.
NOTE
The amplitude calibration must be performed after at least 30 minutes warming-up of the
system.
Followings are the procedures for the amplitude calibration of E4402B Spectrum
Analyzer:
1. Run the VEE Test Environment Software after the warming up the system. On the
Main Menu display, click on [Configuration] to open the Configuration menu.
Figure 3-13
2. On the Configuration Menu display, click on [SA Calibration] button to open the SA
Calibration menu.