3-35
Triggering the Oscilloscope
To use Sequence triggering
5 If you select an edge term, one channel must be set to a rising edge or
a falling edge. All other channel edges will be set to don’t care (X).
a Press the
Channel
softkey (or rotate the Entry knob on mixed-signal
oscilloscopes) to select the channel.
The channel you select is shown in the upper-right corner of the display next
to "
Seq
".
b Then press
the
X
softkey to select rising edge or falling edge. All
other channel edges will be defaulted to don’t care.
6 When you set a Trigger on: condition to trigger on Edge 2, you can also
select which occurrence of Edge 2 to trigger on.
a Make sure
Nth Edge 2
is selected in the
Trigger:
softkey
b Press the
Term
softkey and select
Count (N)
.
c Press the
N
softkey, then turn the Entry knob to set the number of edges to
wait before triggering.
If the Find event occurs again before the Count (N) event is satisfied,
Count (N) will be reset to zero.
N
can be set from 1 to 10,000.
Define the optional Reset on: stage
1 Press the
Stage
softkey and select
Reset on:
.
Reset on: is the last stage in the trigger sequence. When you select the
Stage
Reset on:
softkey, the next softkey to the right will display
Reset:
and give you a
list of the terms you can define for the Reset on: stage. The Reset on: stage can
be set to one of the following conditions:
No Reset
- Resets on the find condition.
Pattern 1
(or
2
)
Entered
– A pattern is entered on the last edge that makes
the pattern true (logical AND).
Pattern 1
(
or 2
)
Exited
– A pattern is exited on the first edge that makes the
pattern false (logic NAND).
Edge 1
(or
2
)
Pattern 1 and Edge 1
Timeout
Terms that are grayed-out are not valid in the reset stage.
Summary of Contents for 54621A
Page 4: ...iv ...
Page 10: ...Contents 6 ...
Page 11: ...1 Getting Started ...
Page 38: ...1 28 ...
Page 39: ...2 Front Panel Overview ...
Page 63: ...3 Triggering the Oscilloscope ...
Page 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 171: ...5 Making Measurements ...
Page 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Page 223: ...6 Utilities ...
Page 239: ...7 Performance Characteristics ...
Page 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Page 260: ...7 22 ...