3- 44
Triggering the Oscilloscope
To use TV triggering
6 If you select a line mode when using standards NTSC, PAL, PAL-M, or
SECAM, press the
Line #
softkey, then rotate the Entry knob to select
the line number on which you want to trigger.
7 If you select a line mode when using the Generic standard, press the
Count #
softkey, then rotate the Entry knob to select the desired count
number.
Listed below are the line (or count) numbers per field for each TV standard.
Line (or count for Generic) numbers per field for each TV standard
TV standard
Field 1
Field 2
Alt Field
NTSC
1 to 263
1 to 262
1 to 262
PAL
1 to 313
314 to 625
1 to 312
PAL-M
1 to 263
264 to 525
1 to 262
SECAM
1 to 313
314 to 625
1 to 312
Generic
1 to 1024
1 to 1024
1 to 1024 (vertical)