3- 34
Triggering the Oscilloscope
To use Sequence triggering
5 If you select an edge term, one channel must be set to a rising edge or
a falling edge. All other channel edges will be set to don’t care (X).
a Press the
Channel
softkey (or rotate the Entry knob on mixed-signal
oscilloscopes) to select the channel.
As you select a channel, the channel is highlighted in the selected pattern
list shown in the waveform area.
b Then press
the
X
softkey to select rising edge or falling edge. All
other channels will be defaulted to don’t care (X).
If you want to reassign an edge to a different channel, repeat the above step.
The value of the original channel edge will be defaulted to
X
(don’t care).
Define the Trigger on: stage
1 Press the
Stage
softkey and select
Trigger on:
.
Trigger on: is the next stage in the trigger sequence. When you select the
Stage
Trigger on:
softkey, the next softkey to the right will display
Trigger:
and give you
a list of the terms you can define for the Trigger on: stage. The Trigger on: stage
can be set to one of the following conditions:
Pattern 2 Entered
– A pattern is entered on the last edge that makes the
pattern true (logical AND).
Pattern 2 Exited
– A pattern is exited on the first edge that makes the pattern
false (logical NAND).
Edge 2
Pattern 2 and Edge 2
Nth Edge 2
2 Press the
Trigger:
softkey and select a stage to trigger on.
3 To define the term(s) used in the Trigger on: stage, press the
Term
softkey and select the pattern or edge term displayed in the
Trigger:
softkey.
4 If you select a pattern term, each channel in the pattern must be set to
a
H
(high),
L
(low), or
X
(don’t care).
a Press the
Channel
softkey (or rotate the Entry knob on mixed-signal
oscilloscopes) to select the channel.
b Press the
L H X
softkey to set a level for the channel.
c Repeat for all channels in the pattern.
Summary of Contents for 54621A
Page 4: ...iv ...
Page 10: ...Contents 6 ...
Page 11: ...1 Getting Started ...
Page 38: ...1 28 ...
Page 39: ...2 Front Panel Overview ...
Page 63: ...3 Triggering the Oscilloscope ...
Page 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 171: ...5 Making Measurements ...
Page 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Page 223: ...6 Utilities ...
Page 239: ...7 Performance Characteristics ...
Page 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Page 260: ...7 22 ...