MegaZoom Concepts and Oscilloscope Operation
To capture a single event
To capture a single event
To capture a single event, you need some previous knowledge of the signal to
set up the trigger level and slope. For example, if the event is derived from TTL
logic, a trigger level of 2 volts should work on a rising edge. These steps show
you how to use the oscilloscope to capture a single event.
1 Connect a signal to the oscilloscope.
2 Set up the trigger.
a Define an edge, pulse width, pattern, or other advanced trigger type.
b Press the
Mode/Coupling
key and set trigger
Mode
to
Normal
.
c If you are using the analog channels to capture the event, turn the trigger
Trigger Level knob to the trigger threshold where you think the trigger
should work.
3 Press the
Single
key.
Pressing the
Single
key arms the trigger circuit and the
Single
key will be
illuminated. When the trigger conditions are met, data appears on the display,
representing the data points that the oscilloscope obtained with one acquisition
and the
Run/Stop
key will be illuminated red. Pressing the
Single
key again
re-arms the trigger circuit and erases the display.
Operating Hints
When Trigger Mode is set to Auto or AutoLvl (auto-single mode), each time
Single
is pressed the oscilloscope will wait for a valid trigger. If no trigger is found after
about 40 ms, the oscilloscope will force a trigger, acquiring whatever signal is
present. This is a handy mode to look for totally unknown situations.
When Trigger mode set to Normal, pressing
Single
will arm the trigger circuit and
do a single acquisition when a valid trigger is seen.
You can use recalled traces to compare several single-shot events. See the "Saving
and Recalling Traces and Setups" section later in this chapter.
Summary of Contents for 54621A
Page 4: ...iv ...
Page 10: ...Contents 6 ...
Page 11: ...1 Getting Started ...
Page 38: ...1 28 ...
Page 39: ...2 Front Panel Overview ...
Page 63: ...3 Triggering the Oscilloscope ...
Page 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 171: ...5 Making Measurements ...
Page 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Page 223: ...6 Utilities ...
Page 239: ...7 Performance Characteristics ...
Page 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Page 260: ...7 22 ...