Performance Tests and Calibration
4
34405A User’s and Service Guide
95
AC Current Gain Adjustment Procedure
Review the
“Test Considerations”
on page 67 and
“Gain Adjustment
Considerations”
on page 90 sections before beginning this procedure.
1
Press
to enter AC Current Gain Calibration.
2
The primary display will show the calibration value and the secondary
display will show the reference value of the
Cal Item
.
3
Configure each
Cal Item
shown in the adjustment table below.
4
Use
(Auto) or
(Range) to select the
Cal Item
.
5
Apply the input signal shown in the Input and Frequency columns of
the table below.
Table 17
AC Voltage Gain Adjustment
Input
Vrms
Frequency
Function
Frequency as 1kHz
Cal Item
10 mV
1kHz
AC Voltage
10 mV
100 mV
1kHz
100 mV
1 V
1kHZ
1 V
10V
1kHz
10 V
100 V
1kHz
100 V
750 V
1kHz
750 V
Caution: Set the calibrator output to 0V before disconnecting from the multimeter input
terminals.
ACI
N O T E
Always complete tests in the same order as shown in the appropriate table.