Performance Tests and Calibration
4
34405A User’s and Service Guide
85
Calibration Process
The following general procedure is the recommended method to complete
a full instrument calibration.
1
Read
“Test Considerations”
on page 67.
2
Perform the verification tests to characterize the instrument (incoming
data).
3
Unsecure the instrument for calibration (see
“Calibration Security”
on
page 82). Once unsecured, the instrument will be in Adjustment Mode
as indicated by the illuminated CAL annunciator.
4
Perform the adjustment procedures (see
“Adjustments”
on page 88).
5
Secure the instrument against calibration.
6
Note the new security code and calibration count in the instrument's
maintenance records.
N O T E
Make sure to quit the Adjustment Mode then turn off the instrument.