36
34405A User’s and Service Guide
2
Features and Functions
Using the Utility Menu
The Utility Menu allows you to customize a number of non- volatile
instrument configurations. It also displays error messages and hardware
revision codes. The contents of the Utility Menu are shown in the table
below.
Primary Display
Secondary
Display Settings
Description
Remote Command
tESt
no
YES
IF YES, immediately execute self-test
upon next Store/Recall button push.
After self-test completes, returns to
normal instrument operation.
*TST? (self-test is executed
immediately)
ºunit
ºC
ºF
Changes displayed units for temperature
measurements
UNIT:TEMPerature <units>
bEEP
On
OFF
Enable, disable Diode, Min/Max, Limit
Test, and Hold beep operations
SYSTem:BEEPer:STATe <mode>
P-On
rESEt
LASt
Enable or disable power-on recall of State
0 (last power-off instrument state). Note:
The multimeter always saves the
power-down state. This just determines
whether or not to recall the state at
power-on.
MEMory:STATe:RECall:AUTO <mode>
2.diSP
On
OFF
Turn the secondary display on or off.
DISPlay:WINDow2[:STATe] <mode>
StorE
On OFF
Enable,
disable
all front panel state store
operation
MEMory:STATe:STORe <mode>
Edit
On
OFF
Enable, disable all math register editing
None
Error
nonE
nn.Err
See Reading Error Messages below.
SYSTem:ERRor?
CodE
1-dd.d
2-dd.d
Displays processor code revision
numbers.
1= Measurement processor revision.
2= IO processor revision.
*IDN? (from remote also returns
manufacturer's name, model number,
and the serial number)
UtitY
donE
Display
donE
on primary display for 1
second then return to normal operation
None
Downloaded from
www.Manualslib.com
manuals search engine