Specifications
6
34405A User’s and Service Guide
129
Supplemental Measurement Specifications
Table 29
Supplemental Measurement
Specifications
DC Voltage
•
Measuring Method:
•
Sigma Delta A-to-D converter
•
Input Resistance:
•
10M
Ω
± 2% range (typical)
•
Input Protection:
•
1000V on all ranges (HI terminal)
Resistance
•
Measurement Method:
•
2-wire Ohms
•
Open-circuit voltage:
•
Limited to < 5V
•
Input Protection:
•
1000V on all ranges (HI terminal)
DC Current
•
Shunt Resistance:
•
0.1
Ω
to 10
Ω
for 10mA to 1.2A ranges
•
0.01
Ω
for 12A range
•
Input Protection:
•
Front Panel 1.25A, 500V fuse for I terminal
•
Internal 15A, 600V fuse for 12A terminal