When you select the Vibration tab, the vibration data stream from 280-DS
Instrument Module starts and other measurement data streams from Vessel
Module and Instrument Module stop. This allows you to stop and restart the
vibration data in order to take a snapshot of the data.
Figure 58.
Vibration Tab
The vibration data can be recorded and replayed. When you click
and
select
the output file, the accelerometer data is saved as the tab-delimited, three
column spreadsheet. Each column represents acceleration on the X-, Y-, and
Z-axis. To stop the recording, click and to see the recorded data, click .
NOTE
You can record up to 60 seconds of the data stream.