220V - 240V / 50-60Hz configuration
Standby
87W
Average continuous scanning
237W
During scanning
590W max.
100V - 120V / 50-60Hz configuration
Standby
92W
Average continuous scanning
245W
During scanning
621W max.
Environmental conditions (during operation)
In line with IEC721-3-2: 2M2 and 2K2 with following restrictions:
Room temperature
Agfa CR HM5.x Mammo: between
+20°Cand+30°C (68 - 86°F)
Other plates and cassettes: between
+15°Cand +30°C (59 - 86°F)
Maximum temperature
change rate
max. 0.5°C/min (0.9°F/min)
Relative humidity
between 15%and75% (non-condensing)
Light
no exposure to direct sunlight, max.
2500Lux
Environmental conditions (during storage)
Room temperature
between -25 °C and 55 °C
Relative humidity
between 10% and 95%
Barometric pressure
between 70 kPa and 106 kPa
Environmental conditions (during transport)
Room temperature
between -25 °C and 55 °C
Relative humidity
maximum 85%
Barometric pressure
minimum 70 kPa
Environmental conditions for mobile installation (during transport)
In line with IEC721-3-5: 5M1 and 5K1 with following restrictions:
82
| DX-G, DX-M | Technical Data