Color Constant/
Flashing
Status
Action
Blue
Constant
Activating the era-
sure cycle.
Enter cassettes into the input
buffer for erasing.
Remove cassettes from the out-
put buffer.
Flashing
Proceeding with the
erasure cycle.
Green Constant
Stand-by mode
(READY).
Flashing
Busy with scanning
and transporting cas-
sette and image
plate.
Enter cassettes into input buffer
for scanning.
Remove cassettes from output
buffer.
Red
Constant
Service mode
Check digitizer touch panel and
workstation display for further
information and detailed in-
structions.
Fatal error
Contact an Agfa certified service
engineer.
Flashing
• Warm up / Self-
test
• Processing Soft-
ware down
• Error
Check digitizer touch panel and
workstation display for further
information and detailed in-
structions.