Prepare the X-Ray system for the examination
Prepare the X-Ray system for the examination:
1.
On the touch screen console, touch the Set button.
CAUTION:
Before initiating automatic movement, check the position of
the floor mount locks of the DX Full Leg Full Spine Stand.
Locks with rubber feet can move out of position when
positioning the patient and cause the stand to collide with
the X-ray system during the examination. If needed, let the
patient step down from the stand and repeat the positioning
of the stand and of the patient.
2.
Press and hold the auto positioning button that moves the X-ray system to
the automatic position.
The X-Ray system moves to the starting position for the examination.
When the position is reached the position status in the Software Console is
changed to ‘on target’.
34
| DR Full Leg Full Spine | Getting started
0179E EN 20210708 1527