When performing a FLFS examination, the system imposes limits to the
bottom and the top of the region of interest, depending on the X-Ray
system installation.
When performing a Full Leg examination, the footsteps must be positioned
to accommodate for the system limitations to the region of interest.
1.
On DX-D 300, the region of interest on the vertical ruler cannot go
lower than ±40 cm.
2.
On DX-D 600, the region of interest on the vertical ruler cannot go
lower than ±25 cm.
3.
On DR 600, the region of interest on the vertical ruler cannot go lower
than ±20 cm.
These values may be slightly different depending on the relative
installation position of the X-Ray system and the DX Full Leg Full Spine
Stand.