4.1.5
RALCO AUTOMATIC COLLIMATOR R225ACS
Collimator controls:
1.
Collimator Display
2.
Manual Blade Control
3.
Automatic Mode Indicator (Green)
4.
System not ready Indicator (Red)
5.
Manual Mode Indicator (Yellow)
6.
Change of Filter
7.
Collimator Lamp Control (Led “ON”)
8.
Retractable Metric Tape
9.
Laser Pointer Window
10.
Laser Pointer On/Off Button
6
2
3
4 5
7
Transverse Field Size
Source to Image Distance
1
Manual Blade Knobs
2
8
9
10
Longitudinal Field Size
1
When pressing the Collimator Lamp push-button, the Collimator light and an
optional Laser light turn on. They remain lighting for 30 seconds before they
switch Off automatically (lighting time can be configured).
Exposure field on the Receptor is automatically adjusted, or when in manual
mode by setting the two knobs. The Collimator Display shows the selected field
size (cross / long).
The Collimator can rotate
±
90
o
on its vertical axis while the Tube remains in the
same position. This movement is performed by manually turning the Collimator
and has detents every 90
o
.
Summary of Contents for 5420/100
Page 1: ...DX D 400 User Manual type 5420 100 Technical Publication 0232F EN 20191213 ...
Page 2: ......
Page 10: ...DX D 400 User Manual vi 0232F EN 20191213 ...
Page 18: ...DX D 400 User Manual 8 0232F EN 20191213 This page intentionally left blank ...
Page 56: ...DX D 400 User Manual 46 0232F EN 20191213 This page intentionally left blank ...
Page 60: ...DX D 400 User Manual 50 0232F EN 20191213 This page intentionally left blank ...
Page 104: ...DX D 400 User Manual 94 0232F EN 20191213 This page intentionally left blank ...
Page 132: ...DX D 400 User Manual 122 0232F EN 20191213 This page intentionally left blank ...
Page 154: ...DX D 400 User Manual 144 0232F EN 20191213 This page intentionally left blank ...
Page 180: ...DX D 400 User Manual Appendix B B 8 APR MATRIX RUSSIAN VERSION ...
Page 187: ......