GUIDANCE AND MANUFACTURER’S DECLARATION -- ELECTROMAGNETIC IMMUNITY
(IEC 60601-1-2:2007)
The DX--D400 radiographic room is intended for use in the electromagnetic environment specified below.
The customer or the user of The DX--D400 radiographic room should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2:2007
Test level
Compliance
level
Electromagnetic environment -- guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
¦
6 kV contact
¦
8 kV air
¦
6 kV
¦
8 kV
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4 (*)
¦
2 kV for power supply lines
¦
1 kV for input/output lines
¦
2 kV
N/A
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000--4--5 (**)
¦
1 kV line(s) to line(s)
¦
2 kV line(s) to earth
¦
1 kV
¦
2 kV
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage
dips,
short
interruptions
and
voltage
variations on power supply
input lines.
IEC 61000--4--11
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in U
T
)
for 250 cycles
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
< 5% U
T
(>95% dip in U
T
)
for 1 cycle
60%
for 5 periods
30%
for 25 periods
>95%
for 250 periods
>95%
for 0.5 periods
>95%
for 1 period
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Radiographic Room requires continued
operation during power mains interruptions, it is
recommended that the Radiographic Room be
powered from an uninterruptible power supply
or a battery.
Power frequency (50 Hz)
magnetic field
IEC 61000--4--8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE -- U
T
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for 5420/100
Page 1: ...DX D 400 User Manual type 5420 100 Technical Publication 0232F EN 20191213 ...
Page 2: ......
Page 10: ...DX D 400 User Manual vi 0232F EN 20191213 ...
Page 18: ...DX D 400 User Manual 8 0232F EN 20191213 This page intentionally left blank ...
Page 56: ...DX D 400 User Manual 46 0232F EN 20191213 This page intentionally left blank ...
Page 60: ...DX D 400 User Manual 50 0232F EN 20191213 This page intentionally left blank ...
Page 104: ...DX D 400 User Manual 94 0232F EN 20191213 This page intentionally left blank ...
Page 132: ...DX D 400 User Manual 122 0232F EN 20191213 This page intentionally left blank ...
Page 154: ...DX D 400 User Manual 144 0232F EN 20191213 This page intentionally left blank ...
Page 180: ...DX D 400 User Manual Appendix B B 8 APR MATRIX RUSSIAN VERSION ...
Page 187: ......