DX-D 400
User Manual
24
0232F EN
20191213
2.7.8
DESIGNATED SIGNIFICANT ZONES OF OCCUPANCY
X-ray equipment specified for examinations that do not need the operator or
staff to be close to the patient during normal use shall be provided with means
to allow the following control functions from a “
Protected Area
” (
refer to
illustration below
):
Selection and control of modes of operation.
Selection of loading factors for the exposure.
Actuation of the exposure controls.
Other necessary controls for the operator during exposure.
EXAM ROOM
Generator
PATIENT ENVIRONMENT
PROTECTED AREA
NO RADIATION AREA
OPERATOR CONTROL